69500:EPSTEIN-BARR VIRUS ANTIBODY TO VIRAL CAPSID ANTIGEN (VCA), IgG | |||||||||
Alias Names: | EBV | ||||||||
Methodology: | Multiplex Flow Immunoassay (Luminex) | ||||||||
Edit Date: | 6/11/2010 | ||||||||
Performed: | Daily | ||||||||
Released: | Same day as tested | ||||||||
CPT Code: | 86665 | ||||||||
Specimen Collection Details | |||||||||
Collection: | One 7.5 mL serum separator tube (SST). Also acceptable: One 5 mL red top tube. | ||||||||
Handling: | Allow to clot 10 minutes. Centrifuge and immediately separate serum from cells into a plastic vial. Refrigerate. Freeze at -20° C if specimen cannot be assayed within 7 days. Allow no more than 3 freeze/thaw cycles. | ||||||||
Stability: | 8 hours ambient, 7 days refrigerated at 2-8° C, or 1 year frozen at -20° C. | ||||||||
Standard Volume: | 1 mL serum. | ||||||||
Minimum Volume: | 0.5 mL serum. | ||||||||
Transport: | Refrigerated if received in laboratory within 7 days of collection, or frozen on dry ice. | ||||||||
Rejection Criteria: | Plasma; gross lipemia; hemolysis; specimen that has been frozen/thawed more than 3 times; heat inactivation; refrigerated specimen received in laboratory more than 7 days of collection, or ambient more than 8 hours of collection. |
Reference Range: | |||||||||||||||||||
<0.8 AI (AI = Antibody Index) EBV IgG Antibody to Viral Capsid Antigen (VCA) Result Interpretation, Luminex Method: | |||||||||||||||||||
<0.8 AI
|
Negative: No significant level of detectable IgG antibody to EBV Viral Capsid Antigen (VCA)
|
||||||||||||||||||
0.9-1.0 AI |
Equivocal: Repeat testing in 10-14 days may be helpful
|
||||||||||||||||||
≥ 1.1 AI
|
Positive: IgG antibody to EBV Viral Capsid Antigen (VCA) detected which may indicate a current or recent
EBV infection |
||||||||||||||||||